Parameter | Description |
Total volume | 0.2L,0.5L,1L,2L,3L,5L,10L,20L,30L,40L etc. |
Working volume | 50%~70% |
Tank | Single glass vessel with stainless steel bottom. Lid made by stainless steel 316L with ports, condenser on the cover |
Stirring system | Mechanical drive(GJ type) or Magnetic drive(GC type, bottom/top drive), 3 six-blade paddle and 1 foam breaker, each paddle height adjustable. Stir speed: 50~1000rpm/50-500rpm |
Sterilization | Off-site sterilization (Autoclave) |
Gas control | Control by rotor meter, air flow 1-2vvm. |
Feeding | Automatically feeding by peristaltic pump |
Anti-foam control | Automatically add antifoamer by peristaltic pump |
Temperature control | Auto-control, electric heating range water source add 5℃ to 70℃ |
PH control | PH 2.00-12.00±0.1, auto-control adding acid/base by peristaltic pump, pH sensor from Hamilton/Mettler, Switzerland. |
DO control | 0-150±3% displaying precision 0.1%, DO sensor from Hamilton/Mettler, Switzerland. DO coupling with agitation, oxygen. |
Sampling | Manual sampling |
Control parameters | 10 or 15 inch touch screen , Siemens PLC with control parameters (stirring, pH, DO, temperature, feeding, alarm, gas etc.), data analysis, data record, data output, language(RU/EN/CN) and other customized functions |
Optional configuration | Remote control, Mass flow meter(auto control gas), O2/N2/CO2, gas(CO2) exhaust analysis, Level sensor, Turbidity(OD) sensor, ORP sensor, weight system, Live cell sensor(Biomass), Printing, Siemens touch screen, etc., |
BLBIO-XGC / XGJ is a set of laboratory glass fermentation equipment, autoclave sterilization off-site, easy to operate, powerful, fermented meet various purposes, such as batch, fed-batch, or animal cell culture . Color touch screen display of all measured values and control parameters, which greatly facilitates the user’s actions. In front of the control box installed four peristaltic pumps, can add acids, bases, defoamers, and medium to the fermentor. Fermentation of various parameters such as, pH, DO, temperature and speed can be monitored by the controller, all the data and control parameters can be passed to the controller or computer. Compact structure, the various components can be integrated on the fermentation system. Using upper driver directly coupled mechanically mixing method, the structure is simple, stable operation.